Wholesale ECI-H8F500 Thermostatic Negative Pressure Follicle Collection Instrument Manufacturers, Suppliers

● Intelligent processor control
● Configuration of wireless foot switch
● Anti-backflow design
● LCD real-time display

Product Description

Introduction

The ECI-H8F500 Ovum Pick-up Equipment uses a dedicated processor CPU to control temperature settings and control negative pressure settings. It has high control accuracy and working efficiency. The LCD screen accurately displays and sets the real-time working temperature and negative pressure values, which makes the data more intuitive and easier to operate. The products are applied to the needs of animal embryo transplantation, cell manipulation and processing, and animal research experiments.

Through the combination of temperature sensor feedback information and PID control algorithm design, the processor realizes the precise control of the temperature of the electric heating system of the thermostat tank, and the real-time temperature can be controlled at the set temperature value with an error of ± 0.5℃.

Through the combination of air pressure sensor feedback information and PID control algorithm design, using PWM control, the processor realizes the precise control of negative pressure on the air circuit system of the brushless negative pressure pump, the real-time negative pressure can be controlled at the set negative pressure value, with an error of ±5MMHg.

Because the processor software switches the negative pressure flow controlled by the solenoid valve accurately and in real time, it avoids the possibility of backflow of oocyte fluid, thus preventing damage to the ovum pick-up instrument and operational failure.

The switch can be configured with a duplex wireless foot pedal. The white pedal corresponds to operation at the normal set-point. The yellow pedal is for boost operation. It is easy to operate and facilitates freeing up both hands for various other operations.

ECI-H8F500 Introduction
Technical Parameters
Working Voltage:
200-240VAC
Working Frequency:
50-60Hz
Rated Power:
120VA
Working Temp:
-20-50℃
Relative Humidity:
20-75%RH
Temp Range:
35.0-40.0℃
Default Temp:
37.0℃
Temp Accuracy:
±0.5℃
Negative Pressure:
0-200mmHg
Max Pressure:
≥500mmHg
Test Tube Wells:
8 Pcs
Weight:
8.38Kg
Size:
310*260*150mm
Display and Operation Panel
Pneumatic connector: Use appropriate hose sizes to connect the stainless steel Pagoda head with the water-blocking filter and the oocyte collection test tube.
Negative pressure setting: Adjustable via knob (0-200mmHg). Clockwise increases pressure, counterclockwise decreases it.
Temperature setting: Range 35.0℃-40.0℃. Use arrow buttons to adjust by 0.1℃ increments. Default is 37.0℃.
Start/pressurization: Hold for >3 seconds during clogging emergencies to trigger maximum pumping force (≥480mmHg).
Liquid crystal display (LCM): Real-time monitoring of set/actual temperature and set/actual negative pressure.
Panel Detail 1
Panel Detail 2
Panel Detail 3
Full Equipment View
Frequently Asked Questions
What is the temperature control accuracy of this equipment?
The equipment utilizes a PID control algorithm and high-precision sensors to maintain temperature with an error margin of only ±0.5℃.
How does the system prevent the backflow of oocyte fluid?
The processor software manages a solenoid valve to switch negative pressure flow accurately in real-time, effectively preventing backflow and protecting the instrument.
What should I do if the pipeline becomes clogged during operation?
You can press and hold the pressurization button for more than 3 seconds. This triggers the maximum pumping force (≥480mmHg) to clear the blockage immediately.
Can the equipment be operated hands-free?
Yes, it can be configured with a duplex wireless foot pedal. The white pedal is for normal operation, and the yellow pedal is for boost mode, allowing for hands-free control.
What are the primary applications for the ECI-H8F500?
It is designed for animal embryo transplantation, cell manipulation, processing, and various animal research experiments requiring precise thermostatic negative pressure.
What information is shown on the LCD display?
The LCD screen provides real-time data for four parameters: the set temperature, actual working temperature, set negative pressure, and actual working negative pressure.